Qubit 8/6/19

E1 E2 E3 E4 E5 P1 P2 P3 P4 P5

Procedure

8119.2 microleters of buffer and 40.8 microleters of quant-it reagent were mixed into a master mixed for 40 reactions (should have been 42) + a 2% error

190 microleters of the master mix was mixed with 10 microleters of standard 1 and standard 2 199 microleters of the master mix was mixed with 1 microleter of each post-PCR DNA sample

The samples were run through the Qubit on the dsDNA - Broad Range setting. The nanograms of DNA per microleter was determinded

Sample 24F #1 24F #2 28S #1 28S #2 IA2 #1 IA2 #2 ITS1 #1 ITS1 #2
E1 45.8 ng/μl 43.0 86.0 83.0 94.0 91.0 31.6 31.0
E2 49.2 ng/μl 48.4 86.0 85.2 105 101 30.2 29.6
E3 53.2 ng/μl 52.6 80.8 78.8 104 103 21.4 21.0
E4 55.2 ng/μl 54.6 82.4 80.8 105 96.0 31.2 30.6
E5 53.2 ng/μl 52.8 80.0 79.2 107 105 29.2 28.8
P1 39.2 ng/μl 38.8 45.2 44.0 19.0 18.6 27.0 26.6
P2 18.2 ng/μl 18.0 26.2 25.8 12.9 12.7 14.7 14.5
P3 39.6 ng/μl 39.4 45.0 44.4 13.2 13.0 10.2 10.1
P4 45.0 ng/μl 44.8 48.0 47.0 17.8 17.5 24.4 24.0
P5 46.2 ng/μl 46.0 8.44 8.20 9.62 9.46 20.8 20.4
Standard 1 165.88 ng/μl              
Standard 2 17385.21 ng/μl              

All of these are in ng/μl

For 24F E1-5 was taken from the 7-15-19 tube. P1-5 was taken from the 7-16-19 tube

For 28S E1-5 was taken from the 7-17-19 tube. P1-5 was taken from the 7-22-19

For IA2 E1-5 was taken from 7-23-19. P1 and P3-5 was taken from 7-24-19 P2 was from 7-29-19.

For ITS1 E1 E2 E3 and E5 was taken from 6-18-19 and E4 was taken from 6-19-19. P1-5 was taken from 6-19-19.

Notes

  • P5 for 28S was weak in the PCR and came out as ~8.3

This Data will be used to create aloquates to send to be sequenced.

Written on August 6, 2019