CA6 PCR #6 8/12/19

E27 E28 E29 E30 E31 E32 E33 E34 E35 E36 E37 E40

Procedure

Samples E27 E28 E29 E30 E31 E32 E33 E34 E35 E36 E37 E40 sed in this PCR (polimerase chain reaction) reaction

First A Master Mix of water, 2x mytaq mix, CA6.38R Primer (concentration: 10 micromolar), CA6.38L Primer (concentration: 10 micromolar)

Component Volume for 14 reactions(+5% error)
Water 132.3 micoleters
2x mytaq mix 183.75 micoleters
CA6.38R Primer (concentration: 10 micromolar) 14.7 micoleters
CA6.38L Primer (concentration: 10 micromolar) 14.7 micoleters
BAS (Borine Serum Albumin) 1 microliter

24 micoleters of this master mix was added to 14 tubes For the 2 negative controls 12.5 micoleters water and 12.5 micoleters mytaq mix were added there was no positive control was available

1 microleter of the DNA samples was added to the coresponding tube

This PCR was set to Previous PCRs were showing up smeared and unclear. This could be becuase of too much amplified DNA

Step Temperature Duration cycles
1 94°C 3:00 1
2 94°C 2:00 1
3 58.2°C 1:30 1
4 72°C 2:00 1
5 94°C 0:30 30
6 58.2°C 1:00 30
7 72°C 0:15 30
8 72°C 2:00 1
9 4°C hold -

these amplified DNA samples were then tested by making a gel

to create the Gel the following steps were taken

1.5g agarose and 75ml 1X TAE buffer were mixed to make a 2% agarose 75ml gel mix the mix was microwaved and swirled until there were no flakes then 1 microliters Gelgreen was added

the mix was left to cool for 5 minutes then poured into the gel mold the gel was left to harden for 30 minutes

once hardened enough Used TAE buffer was poured into the gel box to cover the mold

5 microliter aliquots of DNA samples E27 E28 E29 E30 E31 E32 E33 E34 E35 E36 E37 E40 e two controls were made 1 microliter of 5x loading dye was added to each

First 3 microliters of the HyperLadder 100bp was added to the first well of the mold then the DNA samples were each added to the wells in the following order

Ladder E27 E28 E29 Negative Control E30 E31 E32 Positive control - E7
Ladder E33 E34 E35 Negative Control E36 E37 E40 Positive control - E7

The voltage was set to 100V and run for 60 minutes

This is the resulting Gel

08-12-19_CA6_Gel3 08-12-19_CA6_Gel4

Notes

Written on August 12, 2019